Providing superior temperature testing and conditioning Chambers and Platforms that are
accurate, fast, compact, and efficient

Sigma Quality is
the Ultimate Value

 

M256 Temperature Chamber

Pictures may show optional items
Click on the picture for details

The Model M256 is available in 2 basic configurations, cryogenic and mechanically cooled. A castered stand comes with all 256 chambers.

The M256 is a performance cryogenic chamber with high temperature ramp rates. 

The M256M and M256MM are mechanically cooled chambers that are available with a number of compressor options depending on your cooling rate requirements.

Standard with a Model C controller, Model C4 programmable controller is optional

Models Description
M256 Cryogenic with L-CO2 or L-N2
M256M Single Stage Refrigerated system
M256MM Cascade Refrigerated system

Single Zone Shock Chamber

The M256 chamber can be configured as a high performance single zone shock chamber that meet the requirements of MIL SPEC 202, Method 107 testing, and is able to achieve -170°C.  Configured with the optional Shock package, the chamber features a top mounted C4 controller and maintains a very compact footprint, and is supplied with hinged lift off door.  Click here to learn more
Chamber Size
  Workspace Exterior
Model Volume Feet3
(Liters)
H x W x D
 Inches (Centimeters)
H x W x D
Inches (Centimeters)
M256 5.8
(165)
25 x 25 x 25
(63.5 x 63.5 x 63.5)
58.5* x 30 x 39
(148.6* x 76.2 x 99.1) 
M256M
M256MM
* add 6.5" (16.5 cm) to height for top mounted controller
Operating Ranges °C
High Limit Low Limit
M256 200
[+400 High temp option]
–65 L–CO2
–100 L–N2
–185 option
M256M 200
[+300 High temp option]
–30
–65  L2 –CO2 Boost option
–100  L–N2 Boost option
M256MM 200
[+300 High temp option]
–60
–65  L2 –CO2 Boost option
–100  L–N2 Boost option
Performance °C
(Center 70% of Range)
Cooling rate
°C per minute
Heating rate
°C per minute
M256 19 §   13 §
M256M 2 §
20 (with Boost Option)
M256MM 2.2 §
20 (with Boost Option)
13 §

§  Cooling Rate

Cryogenic rates are highly dependent upon coolant type
(L–N2 or L–C02), pressure, and quality.  For more information download or request a copy of our brochure on Cryogenic Cooling

Mechanical refrigeration (not available on all chambers) rates are dependant on refrigeration system chosen and other factors.
Note: Cryogenic boost can be added to mechanically refrigerated systems to speed transition temperatures

    Heating Rate

Temperatures shown are for typical heaters. 
Faster and slower rates available.

 